SGR - Scintillation Gamma Radiameter


Overview Technical specifications Download

Detector:

  • NaI(Tl), dia 51 x length 51 mm (6 in3)

Sampling:

  • 0.2 s to 120 s, standard or fast response

Measuring range and resolution:

  • up to 250 µGy/h, resolution 0.1 nGy/h
  • up to 250000 cps, resolution 0.1 cps
  • 50 to 3000 keV or 400 to 3000 keV

Alarm:

  • external audio output
  • threshold setting from 50 - 2000 nGy/h
  • variable frequency from 250 to 4000 Hz depending on the measured dose rate (upper frequency corresponds with 10-time multiple of the actual threshold)

Display:

  • large graphical LCD display 320 x 240, white backlight

Power supply:

  • 3 pcs of C cells (single use or rechargeable - NiCd, NiMh)
  • working time 3 - 6 days

Casing:

  • waterproof with increased resistance against mechanical damage and shocks

Dimensions & weight:

  • 290x120x205 mm
  • 2.3 kg

Ambient temperature:

  • -20 to +60 °C


----